Home » Vikane(r) Monitoring Products » GF1900 Sulfuryl fluoride [Vikane®/ ProFume®] Monitor
GF1900 Sulfuryl fluoride [Vikane®/ ProFume®] Monitor

| SPECIFICATIONS |
| Measuring Range | 0-50 ppm full scale |
| Accuracy | ±1 ppm |
| Lag Time | 3 seconds |
| Rise Time | 120 seconds to 90% of final value |
| Fall Time | 30 seconds to 10% of original value |
| Calibration | Against standard gas mixture |
| Pump | Integral, diaphragm type |
| Internal Tubing | 1/4″ OD x 1/8″ ID (6.35 x 3.18 mm) Teflon® |
| Tube Fittings | Compression-type: brass, polypropylene. Hose-barb type: polyethylene |
| Pyrolyzer | Mullite (3Al2O3 • SiO2) housing over Nichrome V element; operating temperature ca. 1000°C (1832°F) |
| Meter | 0-1mA pivot and jewel, 50 divisions full scale |
| Carrying Case | Cordura®/Nylon, provided with 4 ft (1.22m) adjustable shoulder strap; equipped with slot for Power Supply, and accessory pouch. |
| Dimensions | 8.5″ H x 7″ W x 12.75″ D (216 x 178 x 324 mm) |
| Weight | 8.5 lb (3.9 kg) |
| Power | 24 VDC, 5A—furnished by Power Supply or Battery Pack |
| | |